Clear the probe:
Clear the residual gel or stained point on the probe tip by flowing clean water;
Clean the probe tip with little liquid detergent by soft and clean cloth;
Clear the probe tip again by flowing clean water;
Wipe the probe and make it dry by soft and clean cloth.
Caution A:
① Probe can only be used by professionally trained operator;
② Probe can not be sterilized and disinfected by autoclave. If it needs to be used in the sterilized occasion, the probe should be covered by disposable sterilized probe cover.
Caution B:
① Don’t let the probe surface be scratched during use;
② Try to avoid dropping down the probe or knocking the probe;
③ Don’t curve or tightly pull the power cord of the probe;
④ Keep the probe clean and dry;
⑤ Connect or disconnect the probe in power “off” status;
⑥ Never let the probe plug or its near power cord be touched by any liquid.
Errors & Messages |
Solutions |
When starting the system, power indicator light is not lit. | 1)Check the power supply cable where on the rear of equipment whether connected already or not; 2)Check fuse, if find the fuse are burned as power supply sudden change, use spare fuse of the same mode (250V 10A) to replace the fuse. |
When starting the system, keyboard light is lit, monitor has output but no ultrasonic image. | 1)Check the power supply cord of the monitor, whether connected already or not; 2)Check the on-off knob of monitor, whether opened up or not; 3)Check the VGA cable whether connected to the VGA 1 of equipment rear or not. |
When starting the system, the monitor output with format, appears “No probe, please connect probe” dialog box. | Switch off power, and check the connection of probe. |
The video printer cannot print when press the [print] key on the big panel | 1)Check whether the video printer electrified or not; 2)Whether there is printer paper or not in the printer; 3) Check whether the video printer well connected to the system or not. |
Image has interference and the image at high depth is not clear | 1)Move and avoid interference; 2)Use separate 220V power cord; 3)Ground the main unit well; 4)Adjust depth parameter of image. |
Appear “Ultrasound Scanner” dialog box on the screen. | The power supply is not stable. Use a voltage stabilizer. |
The date and time on the screen is not correct. | Press [SETUP] key , select [System] and set the correct time and date. |